Temperature
Range
: -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Temperature Compensation
Automatic, 0 to 100 OC.
User-selectable ATC
coefficient,
0 to -6.0%/OC
Calibration
|
Automatic calibration via
Menu system. |
|
Any Conductivity, TDS, or
Resistivity standard
can be entered for calibration. |
Conductivity Sensor Span Range
?5% of nominal k factor
Temperature Sensor Offset Range
-10.0 OC to +10.0 OC
Memory
2300 readings including date
and time.
Automatic Datalogging
User-set for one reading every
2 to 90 seconds, minutes or hours.
RS232 Port
1200, 9600, 19200 and 38400
Baud.
8 Bits, No Parity, 1 Stop Bit,
XON/XOFF Protocol
Recorder Output (optional)
Conductivity...: 0 to 2000 mV
for all
ranges
TDS............: 0 to 1000 mV
for all
ranges
Resistivity....: 0 to 2000 mV
Ratio..........: 0 to 1500 mV
Manual range selection or
range hold
function should be used to
maintain
logical output.
Output impedance approx 1000
Ohms.
Resolution approx 2 mV.
|
Clock
24 Hour Calendar clock.
Displays date, month, year,
hours, minutes, seconds.
Good Laboratory Practices
Date, time and value of last
successful calibration for each parameter are stored. This information can
be recalled or sent to the RS232 port at any time.
Calibration Alarm
Calibration Alarm can be
activated for parameter of user's choice. Calibration Alarm can be set to
Daily, Weekly, 2 Weekly or 4 Weekly.
Display
240 x 64 dot graphic LCD with
large, bold digits, user-friendly menu system and full-text help and error
messages.
Enclosure
Tough, high impact ABS. Splash
resistant one-piece lid.
Dimensions
240 x 180 x 105 mm
Mass
Instrument only : Approx 0.8 kg
Full Kit
: Approx 3.0 kg
Environment
Temperature : 5 to 45 OC
Humidity : 0
to 90% R.H.
Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified
country is supplied.
|