Conductivity 
       
      Ranges 
       
      (automatic range selection) 
      k=0.1 Sensor : 0 to 2.000 uS/cm 
                                 0 to 20.00 uS/cm 
                                 0 to 200.0 uS/cm 
                                 0 to 2000 uS/cm 
      k=1.0 Sensor : 0 to 20.00 uS/cm 
                                 0 to 200.0 uS/cm 
                                 0 to 2000 uS/cm 
                                 0 to 20.00 mS/cm 
      k=10 Sensor : 0 to 200.0 uS/cm 
                                0 to 2000 uS/cm 
                                0 to 20.00 mS/cm 
                                0 to 200.0 mS/cm 
      Resolution 
         
        0.05% of selected range 
      Accuracy 
         
        ±0.5% of full scale of selected range at 25 ¡É 
         
        TDS 
         
        Ranges 
         
        (automatic range selection) 
        k=0.1 Sensor : 0 to 1.000 ppm 
                                    0 to 10.00 ppm 
                                    0 to 100.0 ppm 
                                    0 to 1000 ppm 
        k=1.0 Sensor : 0 to 10.00 ppm                            
        0 to 100.0 ppm 
                                    0 to 1000 ppm 
                                    0 to 10.00 ppk 
        k=10 Sensor : 0 to 100.0 ppm                           
        0 to 1000 ppm 
                                   0 to 10.00 ppk 
                                   0 to 100.0 ppk 
      Resolution 
         
        0.1% of selected range 
      Accuracy 
         
        ±0.5% of full scale of selected range at 25 ¡É 
         
        pH 
        Ranges : 0 to 14.00 pH 
        Resolution : 0.01 pH 
        Accuracy : ±0.01pH 
         
        mV 
        Ranges : 0 to ±600.0 and 
        0 to ±1500 mV 
        (auto-ranging) 
        Resolution : 0.1 & 1 mV 
        Accuracy : ±0.15 & ±1 mV 
         
        Temperature 
        Ranges : -10.0 to 120.0 ¡É 
        Resolution : 0.1 ¡É 
        Accuracy : ±0.2 ¡É 
         
        Input Impedance 
        >3×1012 Ohms  
       
       
       
     | 
     
      Temperature Compensation 
       
      Conductivity 
       
      Automatic, 0 to 100 ¡É, via thermistor in  
      Conductivity sensor. 
      pH 
         
        Automatic(via Temperature sensor) or manual, 
        0 to 100 ¡É. 
        The labCHEM-CP automatically detects whether or not 
        the temperature sensor is connected. 
         
         
        Automatic Standard Recognition 
         
        Conductivity 
         
        14.94 uS/cm, 73.9 uS/cm, 150.0 uS/cm,  
        717.8 uS/cm, 1413 uS/cm, 2760 uS/cm,  
        6.67 mS/cm, 12.90 mS/cm, 24.82 mS/cm, 
        58.00 mS/cm, 111.9 mS/cm  
      TDS 
         
        69.5 ppm, 2.00 ppk, 8.00 ppk, 36.0 ppk 
      pH 
        4.00, 6.88, 7.00, 9.23 and 10.06 pH 
         
        Enclosure 
        Tough, high impact ABS. Splash resistant one-piece 
        lid. 
         
        Display 
        14.5mm high, 16 character alphanumeric LCD with 
        full-text messages. 
         
        RS232 Output(Optional) 
        300, 1200 and 9600 Baud. 
        8 Bits, No Parity, 1 stop Bit, XON/XOFF  
         
        Recorder Output(Optional) 
        0 to 2000mV Output for¡¦ 
        0.00 to 14.00 pH 
        -1500 to 1500 mV 
        -10.0 to 120.0 ¡É 
        Output impedance approx 1000 Ohms  
         
        Power 
        12V AC/DC Power Adaptor.  
         
        Dimensions 
        270×210 ×75 mm  
         
        Mass 
        Instrument only : Approx 1.0kg 
        Full Kit : Approx 2.5kg  
         
        Environment 
        Temperature : 0 to 45 ¡É  
        Humidity : 0 to 90% R.H.  
         
        Asymmetry Range 
        -1.00 to +1.00 pH  
         
        Slope Range 
        85.0 to 105.0 %  
         
     |